Type in key words to search for your topic interest..

 

 

 

Measurement of ambient noise levels  

ANSI S3.1 -1999   

Measurement instrumentation

HTL measurements above/below 0 dB

Ambient noise levels should be measured annually or whenever a new noise source (i.e. new equipment, new ventilation etc) is introduced in the testing environment.

When conducting ambient noise measures, it is important to do so with all the test equipment turned on.  This way, internal noise that would normally be present during testing will be captured in the measurements.  Conducting measurements without the equipment and lights turned on may deliver low ambient noise levels that are not reflective of the test situation.   

The measurement needs to be conducted in the test room with the microphone of sound level meter located at the place where the listener’s head would normally be situated.  Measurements need to be taken at octave and one-third octave bands (125 – 8000 Hz) at the intervals indicated on table II and III. 

If a sound level meter is not available, sound measurement can be done using a psychoacoustic check method.  This should only be used when the tests conducted in the sound booth use ears covered.  In a psychoacoustic check, two normal listeners are needed.  Each listener will be placed in the test position with earphones on, as if he was going to be tested.  The audiometer dial is then set to 0 dB for each test frequency and the listener is required to respond to each test frequency at this level.  What each listener is required to do is, listen for beats, ambient noise masking and direct acoustic radiation from the audiometer or any other instrument present in the environment.  The listener is also required to check for the presence of intense low frequency vibrations that may be present in the test room.  This is to ensure that low frequency vibrations are not transmitted to the ear through bone conduction.  Low frequency vibration can be transmitted if the listener touches a table or headrest or some other fixed object in the test room.  Low frequency checks are performed with the audiometer turned off.  According to ANSI S3.1 – 1999, the test room is satisfactory for testing hearing at RETSPL if each of the normal listeners responds to each test frequency pure tone at 0 dB without reporting these above mentioned problems.

 

 

Note:  This site is best viewed using IE 4.X or higher under 800x600 resolution


Michael S.H. Der 
Copyright © 1999 Homeboy Homework Co. 
All rights reserved. Revised: April 13, 2001   

click here to email me     

line   line


The current time is EST

 

Audiometer Topics

Bone Conduction Topics

Sound Field Topics

Immittance Topics

Other Topics


 

Audiometers

Bone Conduction

Immittance

Miscellaneous

 


The University of Western Ontario

School of Communication Sciences and Disorders

 

 

line

Disclaimer | Contact Us | FAQ | About Me! | Links

 

TOTAL NUMBER OF VISITORS