Room Construction and Characteristics
Common problems with
sound booths
"Construction
Safety Factor"
Moving a sound booth
The
most accurate sound field test measurements are those which are conducted
in an anechoic
room. In an anechoic testing room, the SPL varies smoothly and will
vary gradually as the distance from the speaker is increased (Walker et al
1984). Unfortunately, the use of anechoic rooms are not cost effective in
a clinical setting and most sound rooms will contain some reverberation.
Because of this less than perfect testing environment, it is important that Audiologists understand some of the limitations and
do's and don'ts regarding their sound
booths.
A study by Siegenthaler (1981) and Frank
and Williams (1993), found very few hearing test booths in clinical use, meeting
the ANSI S3.1* ears covered
standard. Audiologists
can help reduce the chances of measurement variability if they adhere to the
following simple steps. Audiologists
need to be aware that practically all attenuation measurements reported by the
manufacturers of hearing test booth, are taken in a controlled laboratory
environment. These controlled
environments are rarely encountered in the clinical setting and this is a reason
why manufacturer specs may differ from clinical results.
To assure compliance with current standards, it is important that a careful acoustical evaluation of the clinic environment (where the
audiometric room will be installed) be performed. This evaluation needs to
be done by qualified professionals
using proper instrumentation before the booth is actually installed.
Some
of the more common areas of problems in tests booths are the door seals, door
fit and ventilation fans. It is
important that these parts of the booth are thoroughly maintained.
The greatest cause of ambient noise in a test booth seems to be caused by
the ventilating fans (Lankford et al 1980).
A study by Lankford et al. 1980 found that every booth tested which was
over 5 years old had faulty door seals. Simple
maintenance and modifications have proved to reduce background noise in the room
and helped improve the noise attenuating characteristics of the booth.
In a study by Frank and Williams (1993), 136 test rooms were evaluated,
revealing that almost all test rooms contained excessive ambient noise at the
lower frequencies (125-500 Hz) caused by the ventilation system.
They also found that most Audiologists knew that the ventilation system
caused much of the ambient noise, and therefore conducted their ambient noise measurements
with the fans turned off. This is
contrary to ANSI standards, which requires that fans be turned on (so as to best
capture the typical test setting).
In
the Frank and Williams (1993) study, they found that replacing the door seals
and/or tightening the door latch significantly lower the test room ambient noise
SPLs by as much as 10 to 12 dB in the lower frequencies (for rooms that did not
pass ANSI standards). Audiologists
need to be aware of the likely problem areas in their test booths so that they
can present questions or constructive comments to those professionals who
calibrate and help maintain their booths.
Even
if all the necessary steps are taken to ensure that the sound booth
adheres to
manufacturer specs, Audiologists need to be aware that ambient noise levels are
rarely steady and tend to fluctuate and increase over time.
Where a single walled test booth sufficed in previous years, a double
walled equivalent may be required to deal with the changing environment.
Trained professionals should check test booths twice a year or
at a minimum, annually.
An
Audiologist thinking of buying a new test booth should purchase a booth with
enough noise reduction to achieve a noise measurement that is 10 dB less than the ears-open specification of ANSI S3.1-1991.
The reason for this is to account for the fact that real people are
installing the sound booth. This is know as the "Construction Safety
Factor". The Construction safety factor is a 10 dB window
which accounts for the effects of adding doors,
ventilation, windows etc… to the sound booth.
Keep in mind that the manufacturer specification refers only to
the attenuation characteristics of the wall and not the actual sound booth.
Therefore, the attenuation characteristics for a 2.5” wall is actually
closer to 10 dB less than the manufacturer specification.
Audiologists
should also have the supplier verify the noise reduction characteristics of the
booth once it is installed. In
addition, the ambient noise levels of the booth with the ventilation turned on
and off should be measured and meet the ANSI S3.1 - 1999 ears-open criterion.
Relocation
of a sound booth should be avoided if possible. When a sound booth is disassembled and reassembled, it is
highly unlikely it will produce the same attenuation characteristics as it did
before. A major concern in taking a
room apart is the bending of the walls. When
a room is initially assembled, acoustic grade caulking is used in the
construction process. This caulking
must be completely removed, walls properly cleaned and then replaced with new
caulking when reassembled in the new location.
If you use a contractor to assemble your room who has never installed a
sound booth before, then it might be wise to use a “Construction Safety
Factor” closer to 15 dB.
*
The ANSI 3.1 Maximum Permissible Ambient Noise Levels for Audiometric Test Rooms
(MPANLs) standard has since been updated to ANSI S3.1 – 1999.
Note: This site is
best viewed using IE 4.X or higher under 800x600
resolution
|