Audiometer Frequency Tolerance and
Transducer Response Curve:
How each may contribute to measurement
variability
The ANSI S3.6 – 1996 Specification for
Audiometers states that the frequency accuracy for each test signal generated by
a fixed frequency audiometer be +/- 1% for a type 1 audiometer, +/- 2% for a
type 2 and +/- 3% for a type 3, 4 and 5.
What this means is that when
you conduct a pure tone test at a particular frequency (e.g.
1000 Hz), the current ANSI standard allows up to a
1% variation from that frequency (e.g. 1000 Hz +/- 10 Hz) for results to be acceptable for type 1
devices.
However, when conducting tests at 6000 Hz using TDH39 earphones, the
permissible tolerance level can and may introduce errors that can seriously
affect pure tone results.
A study by Guelke et al (1977) was
conducted to see the effect of frequency tolerance on audiometer accuracy. They measured the frequency response characteristic of the
TDH39 receiver on a type 9A coupler and found a sharp resonance peak at 6000 Hz.
Below is a copy of the frequency response curve obtained by Guelke et al
(1977). Note the exceptional peak
in and around the 6000 Hz area.

One point to keep in mind is that in 1977
the current tolerance according to ANSI S3.6 was 3%.
It was not until 1996 that a reduction of tolerances to +/- 1% (type 1
audiometers) took place
in the ANSI S3.6 revision.
This resonant peak in the TDH39 response
curve can introduce a significant amount of measurement variability into pure
tone testing. Because of this
resonant peak, any variation around 6000 Hz can cause a significant variation in
the dB value from the audiometer. A
standard deviation of 6.2 dB at 6000 Hz has been reported in Guelke et al
(1977). Because the resonance point
of the TDH39 earphone is likely to be close to 6000 Hz, it is probable that the precise frequency of
the audiometer will affect pure tone results (as well as calibration).
Guelke et al (1977) found that with the current frequency tolerance
standard (+/- 3% at that time), a variation of 7 dB at 6000 Hz could occur.
This is to say, if the frequency emitted by the audiometer at 6000 Hz was
actually +/- 3% different (i.e. 5820 - 6180 Hz), then a 7 dB difference in SPL
can occur due to the resonant peak of the TDH39 receiver.
Audiologists who have their audiometer
calibrated using TDH39 earphones need to be aware that a harmonic distortion
tolerance of +/- 3% at 6000 Hz may not be sufficient.
An alternative would be to permit only type 1 audiometers to be used for
diagnostic purpose or select a more
stable (flatter response) transducer during the calibration process.
Guelke at al (1977) suggested using a new
receiver, the TDH50 be used in place of the TDH39.
The advantage of the TDH50 receiver is that the resonance at 6000 Hz is
damped (not as sharp as the TDH39). With the damped
resonant peak it is no longer necessary to increase the accuracy of the
frequency output in the audiometer. That
is, it is less important with the TDH50 that the audiometer frequency at 6000
Hz, be actually 6000 Hz.
Audiologists
who have their audiometers calibrated to ANSI S3.6 1989
specification or who are using audiometers classified as
type 3 or greater may be allowing measurement variability
to enter their test results at 6000 Hz, despite having
what they feel to be properly calibrated
equipment.
The take home message here is that
Audiologists need to be aware of the frequency response characteristic of the
transducer they decide to use. Even though their audiometer may meet the current
standard (ANSI S3.6 - 1996), it does not necessarily mean that variation in the signal
output will not occur, as indicated in Guelke et al (1977).
For
topics related to the calibration of TDH 39 earphones,
click on the link below.
Calibration
of the TDH 39 earphone
Note: This site is
best viewed using IE 4.X or higher under 800x600
resolution
|