Calibration variation using the
Radioear B-71 and B-72 Bone Vibrators
Audiologists need to be aware that using a +/-3% frequency
tolerance for generating and calibrating reference bone conduction thresholds is
prone to variability. To minimize
test variability for bone conduction output during audiometer calibration, a
test signal that is as close to the center frequency as possible is desired.
Audiometer manufacturers are now including the Radioear
B-71 or B-72 bone vibrators as standard equipment.
The B-72 is larger than the B-71 because it contains an additional
dynamic mass, which lowers its resonances compared to the B-71.
The graph below was obtained from a study by Richards et al. 1982.
It indicates the frequency response of both instruments.

From Richard et al. 1982
From the graph three resonant peaks can be seen for both
instruments. Due to the added
dynamic mass present on the B-72, the resonant peaks occur at lower frequencies
than the B-71. The resonant
frequency of the B-72 vibrator is approximately 250 Hz, which is slightly lower
than the B-71. Because of this lower resonant frequency, the output is
great for the B-72 at 250 Hz compared to the B-71.
This characteristic
of the B-72 allows it to be operated at higher output levels with less
distortion (greater dynamic range) compared with the B-71 at the lower frequencies.
When calibrating the bone transducer it is important to
know the frequency response of the instrument and more importantly the
characteristics of the low and high frequency slope of the resonances.
As a general rule, when any transducer is operated on or near a steep
resonant slope, the repeatability of measurements at that frequency decreases.
A study by Richard’s et al. tested the repeatability of measurements
for both the B-71 and B-72 by holding the input voltage constant at 0.1 V-rms
and monitoring the output voltage of the artificial mastoid (Bruel and Kjaer,
model 4930) while audiometric center frequencies were varied by +/-3%. The results for this test are indicated in the table below
(from Richards et al. 1982). Notice
that for all test frequencies, the output voltage was less than or equal to 2 dB
except for the B-72 at 4 kHz which was 6.9 dB.

from
Richard et al. 1982
This extreme value at 4 kHz for the B-72 is consistent with
the fact that the frequency response curve for the B-72 near 4 kHz shows a steep
slope.
The study by Richards et al. (1982) used an output range
which was varied over +/-3% at each audiometric frequency. +/-3% was done within the frequency tolerance limits for a
pure-tone audiometer according to the American National Standards Institute
1969. American National Standards
Specifications for Audiometers. ANSI
S3.6-1969 New York. Since the
publication of this study, the tolerances have been made more stringent and
changed from +/-3% to +/-1% for type 1 audiometers and
+/-2% for type 2 audiometers. Audiologists need to be
aware of this revision when their equipment is being calibrated.
In
summary
Audiologists must be aware that using a +/-3% frequency
tolerance for generating and calibrating reference bone conduction thresholds
can produce variability in test retest results. A bone vibrator with a flatter frequency response is
desirable for test retest consistency but flattening out the response curve
would sacrifice the output capability of the instrument. Since this is clinically not desirable, Audiologists must keep
in mind that it is very important during audiometric calibration to have the
test frequency as close to the center frequency.
By doing so, the bone conduction output variations that could influence
calibration and testing can be minimized.
Note: This site is
best viewed using IE 4.X or higher under 800x600
resolution
|