Temperature effects on mechanical couplers
Audiologists
should know that the performance of the mechanical couplers used to calibrate
bone-conduction devices is susceptible to temperature effects.
The effects of temperature are often overlooked but Audiologists need to
keep in mind that errors calibrating audiometers are directly transferred onto
the audiograms of patients. The
mechanical coupler system is often transported from place to place by the person
performing the calibration and depending on the outside and inside temperature
of various clinics, errors in calibration can occur.
Current
IEC standards requires a temperature of 23oC +/- 1 when using a
mechanical coupler. A popular
mechanical coupler system is the Bruel and Kjaer model 4930 shown below.
click to enlarge
When
the temperature of the coupler is different from that stated in the standard, a
correction value cannot be applied to measurements performed at
these temperatures. The reason for
this is because a change in the temperature will affect both the force
sensitivity of the device and its mechanical impedance.
When the mechanical impedance is altered, this affects the force output
of the bone vibrator.
Dowson and McNeill
(1992) conducted a study on the effects of temperature on
the performance of audiometric mechanical coupler systems.
They conducted a study to
measure the performance of a mechanical coupler in temperatures varying from 18
– 33 degrees C. In addition, they
also measured the ability of the coupler to reach a room temperature of 23
degrees C from 5 degrees C.
Dowson
and McNeill (1992) found that measurements of force sensitivity level and
mechanical impedance level generally gave standard deviations of less than 0.2
dB with isolated cases reaching as high as 0.6 dB. Of the three couplers tested, all three showed a significant
temperature effect at higher frequencies (4 & 5 kHz).
They stated that if measurements were made at 18 degrees C instead of 23
for the test couplers, then an error in calibration approaching 3 dB would occur
at higher frequencies due solely to temperature effects.
Dowson
and McNeil (1992) also found a temperature dependence on voltage output when
driven by a bone vibrator type B-71. They
gave an example whereby if a mechanical coupler was stored overnight in 13
degree C temperature and then coupled to a bone vibrator to calibrate an
audiometer, an error of 5 dB would result at the higher frequencies based solely
on temperature effects. Audiologists need to know that if this happens, then this
error gets built into the audiometer’s calibration.
These results may not be typical of all coupler and bone vibrator
combinations.
The
thermal capacity of a mechanical coupler is used to determine how long it takes
for it to reach the desired standard calibration temperature when previously
stored at another temperature.
Dowson and McNeil’s (1992) test couplers required 3 hours to reach 20
degrees C from 5 degrees C, and a further 1.6 hours to reach 22 degrees C.
It is important to
keep in mind that with the exception of extreme cases, the effect of temperature on the
mechanical coupler often may not be large enough to cause
calibration errors resulting in clinically significant
threshold differences. What is important is that
these effects get built into the audiometer's calibration
and when combined with other possible errors in
measurement variability, can result in an additive amount
of error that can have significant effects on clinical
results.
The
take home message here is that a simple element such as temperature can affect the calibration of your audiometer. Since it is not
something you can physically see it can be easily overlooked by both you and the
person calibrating your audiometer.
Note: This site is
best viewed using IE 4.X or higher under 800x600
resolution
|